Add your logo, state the reason for the comparison, record the date, and identify both laboratories. The cover page turns a worksheet into a report your document control system can hold.
Free Excel Template
No proficiency testing scheme covers your measurement, and your assessor still wants evidence that your results hold up. This interlaboratory comparison template turns two sets of results into a single pass or fail decision. Enter both laboratories' values and expanded uncertainties, and the workbook calculates the normalized error at every point.
Tell us where to send it, and the download appears right after you submit the form.
The workbook holds the two documents an assessor expects to see: a cover page that explains why you ran the comparison, and a data page that shows the arithmetic behind your conclusion.
Add your logo, state the reason for the comparison, record the date, and identify both laboratories. The cover page turns a worksheet into a report your document control system can hold.
Line up both laboratories side by side, matched by nominal value. Each row carries the two results and the two expanded uncertainties for one measurement point.
The normalized error formula sits in the sheet. Enter the four inputs for a point, and the workbook returns the En value without a side calculation.
Each point reports against the |En| ≤ 1 rule, so a reviewer sees the outcome without interpreting the number. Failures stand out for investigation instead of hiding in a column.
Capture the coverage factor each laboratory used. Mixing a k = 2 uncertainty with a k = 1 uncertainty distorts the comparison, so the sheet keeps that detail visible.
Print the cover page and the data summary, attach both calibration certificates, and the evidence package is complete. Assessors follow it without asking you to fill gaps.
The numerator holds the raw difference. The denominator holds the difference the combined uncertainty already allows. A point passes when the absolute value of En stays at 1 or below.
The sequence holds across measurement disciplines. Work through it in order, because the first step protects everything after it.
Accreditation bodies generally expect proficiency testing first. State your reason and secure approval before you rely on a comparison.
Pick a device or sample with a current certificate or reported result from an accredited laboratory, covering the points in your scope.
Run your own calibration or test on that item. Capture data at each nominal point you plan to compare.
Enter both readings by nominal value, record each expanded uncertainty, and confirm both laboratories report at the same coverage factor.
Check every matched point against the |En| ≤ 1 rule. Investigate and document any point that fails rather than repeating the measurement until it passes.
Combine the cover page, the data summary, and both certificates into one report, and file it with your validity monitoring records.
Fill in the short form below, and the download unlocks as soon as you submit. We use your details to send the file and occasional ISO 17025 guidance. You can unsubscribe at any time.
This excel file will help you do the following:
Two laboratories received portions of the same water sample and determined three analytes by ion chromatography. Both reported results in milligrams per liter with expanded uncertainties on the same coverage basis.
| Analyte | Lab 1 result | Lab 1 U | Lab 2 result | Lab 2 U | Difference | Combined U | En | Evaluation |
|---|---|---|---|---|---|---|---|---|
| Fluoride | 1.02 | 0.06 | 0.99 | 0.05 | 0.03 | 0.078 | 0.38 | Satisfactory |
| Chloride | 25.4 | 0.80 | 25.0 | 0.70 | 0.40 | 1.063 | 0.38 | Satisfactory |
| Bromide | 0.48 | 0.04 | 0.50 | 0.05 | −0.02 | 0.064 | −0.31 | Satisfactory |
Every point sits well inside the acceptance rule, so the comparison supports the method. Calibration laboratories use the same layout, with nominal load points in place of analytes.
The workbook fits any laboratory that has to show its results agree with someone else's. Use it for a single new parameter or across a whole scope.
No provider covers your measurement, or the available scheme stops short of your range. A comparison gives you evidence that meets the same requirement.
You expanded your scope and the next proficiency testing round sits months away. Use a comparison for interim verification, and document it properly.
You need the calculation, the decision, and the certificates in one package. The workbook gives you a layout you can reuse for every verification.
ISO/IEC 17025:2017 clause 7.7 asks you to monitor the validity of your results. It points to comparison with other laboratories where that comparison is available and appropriate. Proficiency testing is the common route, and an interlaboratory comparison serves the same purpose when a scheme does not fit.
Accreditation bodies generally expect proficiency testing as the first course of action. State your reason, and get your monitoring plan approved before you rely on a comparison. Verify clause references against your controlled copy of the standard.
A point passes when the absolute value of En stays at 1 or below. The statistic divides the difference between the two results by their combined expanded uncertainty. A value at or under 1 means the gap sits inside the uncertainty both laboratories already declared.
The normalized error is the accepted agreement statistic for proficiency testing and comparison work. Its formal definition sits in the statistical standards for those programs. Confirm the exact definition and any coverage factor requirement before you rely on it.
Standardize before you compare. Mixing a k = 2 uncertainty with a k = 1 uncertainty distorts the result. The denominator no longer represents the same confidence on both sides. When the other laboratory uses a different coverage factor, expand your own uncertainty to match it, and record the factor you used.
Both. A calibration laboratory compares readings on a shared artifact at matched nominal points. A testing laboratory analyzes portions of the same sample and compares reported results for each analyte. The arithmetic does not change, and the workbook handles either layout.
Investigate it. A failed point is data, not a formatting problem, so do not repeat the measurement until it passes. Run it through your nonconforming work process and your corrective action process, and record what you found and what you changed. A poor result handled well demonstrates a functioning management system.
No. Treat the comparison as the option you use when proficiency testing is unavailable, out of range, or disproportionate to the verification. Your accreditation body sets the expectations for your monitoring program, and those expectations usually put proficiency testing first.
Your side comes from your own uncertainty budget for the method, and the other laboratory's comes from its certificate or report. If you have not built a budget for the method yet, start there. The En calculation only carries as much weight as the uncertainty estimates behind it.
The comparison depends on sound uncertainty estimates. These resources cover the work that feeds it.
Build the expanded uncertainty you enter on your side of the comparison, with a cover sheet, Type A data, and a contributor table.
Get the template →Four steps from measurand to expanded uncertainty, with a worked example and a column by column walk through the budget.
Read the guide →Guides, technical articles, and templates for testing and calibration laboratories, updated as we publish.
Browse the library →Precision ISO helps laboratories plan validity monitoring and build the uncertainty budgets the En calculation depends on. We also help assemble evidence packages that hold up during an assessment.
This template is a customizable starting point. Review and adapt it to your organization's scope, methods, and processes before use. It does not constitute legal, regulatory, or accreditation advice, and it guarantees no assessment outcome. Clause references point to ISO/IEC 17025:2017, clause 7.7, and uncertainty terms follow the GUM (JCGM 100:2008). Verify every clause reference, the normalized error definition, and any coverage factor requirement against your controlled copies of the standards.
Get practical ISO 17025 guidance sent monthly—no fluff, just what helps labs get accredited
Get practical ISO 17025 guidance sent monthly—no fluff, just what helps labs get accredited